Accuracy of Wearable Devices for Measuring Heart Rate During Conventional and Nordic Walking
Sora Baek1,2, Yuncheol Ha1, Hee-Won Park1,2
1Department of Rehabilitation Medicine, Kangwon National University Hospital, Chuncheon, South Korea.
Introduction:
Nordic walking is being used increasingly as an exercise method in many clinical disorders. To apply Nordic walking in cases of fragile or deconditioned patients, monitoring of exercise intensity such as heart rate (HR) measurement is required. The accuracy of wearable HR monitors during Nordic walking has not yet been reported.
Objective:
To compare the accuracy of an electrocardiography (ECG)-based HR monitor (Polar H7) and a photoplethysmography (PPG)-based HR monitor (Fitbit Charge 2) during conventional and Nordic walking.
Design:
Accuracy was assessed by comparing the HR values obtained using the wearable devices with those obtained via 12-lead ECG as a reference.
Setting:
Laboratory setting.
Participants:
Fifteen male volunteers age 23.7 ± 3.0 years.
Interventions:
None.
Main Outcomes Measures:
HR was simultaneously recorded via 12-lead ECG, the Polar H7, and the Fitbit Charge 2 during conventional and Nordic walking. Agreement between the devices was assessed by calculating Lin's concordance correlation coefficient (rc ), the mean absolute difference, and the limit of agreement (LoA) from Bland-Altman plots.
Results:
Regarding HR values including Nordic and conventional walking, there was a better agreement between the Polar H7 and the reference (rc = 0.96) than between the Fitbit Charge 2 and the reference (rc = 0.84). For the Polar H7, the mean absolute difference from the reference did not differ significantly between the walking methods; for the Fitbit Charge 2, the mean absolute difference was significantly higher during Nordic walking than during conventional walking (6.60 vs. 3.68 bpm, P < .001). The Fitbit Charge 2 had a wider LoA than did the Polar H7 during both walking methods.
Conclusion:
ECG-based wearable devices may be better than PPG-based devices for monitoring HR during Nordic walking. However, both types of devices may adequately monitor HR during conventional walking.
Related Concept Videos
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...