Related Experiment Video
Updated: Aug 1, 2026

Bile Duct Ligation in Mice: Induction of Inflammatory Liver Injury and Fibrosis by Obstructive Cholestasis
Published on: February 10, 2015
The degradable performance of bile-duct stent based on a continuum damage model: A finite element analysis
Xiaoyu Ni1, Yanhong Zhang1, Changwang Pan2
1School of Mechanical and Electronic Engineering, Nanjing Forestry University, Nanjing, China.
Abstract:
Biomedical magnesium alloy stents have become a hot bed of research focus in interventional therapy for nonvascular diseases. In this study, a numerical model for a balloon-expandable bile duct stent made of magnesium alloy with laser sculpture is developed to predict the effects of the degradation of the stent on the biomechanical behavior in the stent-bile duct coupling system. Based on a continuum damage model, the degradable model of the stent is built to understand its performance in an idealized bile duct as it is subject to corrosion over time. The degradation model developed in this study addresses the uniform corrosion and pitting corrosion. By means of the secondary development function of commercial numerical software ANSYS, the finite element analysis procedures were written to control the degradation process based on the technology of element "birth and death," and it is shown how the three-dimensional model and approach give the possibility of analyzing for the degradation mechanism of a magnesium alloy stent in the bile duct or other nonvascular cavities.
Related Concept Videos
Stress-Strain Diagram - Ductile Materials
Yield Criteria for Ductile Materials under Plane Stress
The Maximum Shearing Stress Criterion, also known as the...