WATCHMAN versus ACP or Amulet devices for left atrial appendage occlusion: a sub-analysis of the multicentre LAARGE

Jakob Ledwoch1, Jennifer Franke, Ibrahim Akin

  • 1CardioVascular Center CVC, Frankfurt, Germany.

Insights

This study compared two left atrial appendage closure devices, WATCHMAN and AMPLATZER Cardiac Plug/Amulet, finding comparable safety and effectiveness in real-world practice. Both devices demonstrated high success rates for preventing stroke and death one year post-procedure.

Area of Science:

  • Cardiology
  • Medical Devices
  • Clinical Practice

Background:

  • Left atrial appendage (LAA) closure is crucial for stroke prevention in patients with atrial fibrillation.
  • Existing LAA closure devices have varying designs, necessitating real-world effectiveness comparisons.
  • Generalizing trial results from one device to another is not always appropriate due to design differences.

Purpose of the Study:

  • To compare the clinical performance of two contemporary LAA closure devices: WATCHMAN and AMPLATZER Cardiac Plug (ACP)/Amulet.
  • To assess procedural success, safety, and one-year effectiveness of these devices in a real-world setting.
  • To provide data for informed clinical decision-making regarding LAA closure strategies.

Main Methods:

  • A non-randomized sub-analysis of the prospective, multicentre Left-Atrium-Appendage Occluder Register - GErmany (LAARGE) registry.
  • Inclusion of 641 patients across 38 centres, divided into WATCHMAN (n=278) and ACP/Amulet (n=340) groups.
  • Assessment of technical success, procedural safety, and Kaplan-Meier estimated one-year composite of death or stroke.

Main Results:

  • High technical success rates were observed for both devices (96% WATCHMAN vs. 99% ACP/Amulet; p=0.007).
  • Procedural safety was comparable between the groups (98% WATCHMAN vs. 97% ACP/Amulet; p=0.55).
  • One-year composite rates of death or stroke were similar (12.0% WATCHMAN vs. 12.9% ACP/Amulet; p=0.79).

Conclusions:

  • Both WATCHMAN and ACP/Amulet occluders demonstrate excellent procedural outcomes.
  • Comparable implantation success and procedural safety were achieved with both devices.
  • No significant differences in long-term effectiveness (death or stroke) were observed between the WATCHMAN and ACP/Amulet occluders.