Dose calculation errors as a component of failing IROC lung and spine phantom irradiations
Sharbacha S Edward1,2,3, Mallory C Glenn1,2,3, Christine B Peterson1,4
1The University of Texas MD Anderson Cancer Center UT Health Graduate School of Biomedical Sciences, Houston, TX, 77030, USA.
Purpose:
Between July 2013 and August 2019, 22% of the imaging and radiation oncology core (IROC) spine, and 15% of the moving lung phantom irradiations have failed to meet established acceptability criteria. The spine phantom simulates a highly modulated stereotactic body radiation therapy (SBRT) case, whereas the lung phantom represents a low-to-none modulation moving target case. In this study, we assessed the contribution of dose calculation errors to these phantom results and evaluated their effects on failure rates.
Methods:
We evaluated dose calculation errors by comparing the calculation accuracy of various institutions' treatment planning systems (TPSs) vs IROC-Houston's previously established independent dose recalculation system (DRS). Each calculation was compared with the measured dose actually delivered to the phantom; cases in which the recalculation was more accurate were interpreted as a deficiency in the institution's TPS. A total of 258 phantom irradiation plans (172 lung and 86 spine) were recomputed.
Results:
Overall, the DRS performed better than the TPSs in 47% of the spine phantom cases. However, the DRS was more accurate in 93% of failing spine phantom cases (with an average improvement of 2.35%), indicating a deficiency in the institution's treatment planning system. Deficiencies in dose calculation accounted for 60% of the overall discrepancy between measured and planned doses among spine phantoms. In contrast, lung phantom DRS calculations were more accurate in only 35% and 42% of all and failing lung phantom cases respectively, indicating that dose calculation errors were not substantially present. These errors accounted for only 30% of the overall discrepancy between measured and planned doses.
Conclusions:
Dose calculation errors are common and substantial in IROC spine phantom irradiations, highlighting a major failure mode in this phantom and in clinical treatment management of these cases. In contrast, dose calculation accuracy had only a minimal contribution to failing lung phantom results, indicating that other failure modes drive problems with this phantom and similar clinical treatments.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Biological Effects of Radiation