Interatrial conduction time from electrograms of cardiac resynchronization therapy devices with left ventricular
Leonardo Marinaccio1, Daniele Giacopelli2, Margherita Resta1
1Department of Cardiology, Presidio Ospedaliero di Piove di Sacco, Piove di Sacco, Italy.
Background:
The interatrial conduction time (IACT) is relevant for atrioventricular delay optimization in cardiac resynchronization therapy (CRT) devices. However, this information cannot be easily used as it requires invasive measurements. We tested whether electrical activation of left atrium (LA) could be detected in CRT devices with left ventricular (LV) sensing and used to estimate IACT.
Methods:
The presence of LA activation on LV channel was evaluated in consecutive patients implanted with CRT and quadripolar LV leads whose sensing was temporarily set in the most proximal polarity (Ring LV4-housing). Estimates of IACT during sinus rhythm and atrial pacing were measured and compared with the values obtained with invasive catheterization of coronary sinus.
Results:
Among six patients (50% female; mean age 73.3 ± 4.9 years) included in the analysis, four (66%) had a visible LA signal on the LV channel. The mean IACT measured with device electrograms was 71 ± 8 ms and 133 ± 15 ms during sinus rhythm and atrial pacing, respectively. These values were equivalent to the measurements obtained during invasive catheterization. Both patients without evidence of LA activation had an LV lead with a short total interelectrodes distance (46 mm) resulting in a significant anatomical distance between Ring LV4 and LA.
Conclusions:
In CRT devices with left ventricular sensing, LA signal could be detected and used to estimate IACT especially if long-spaced electrodes are used.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:40Impact of Intracardiac Neurons on Cardiac Electrophysiology and Arrhythmogenesis in an Ex Vivo Langendorff System
Published on: May 22, 2018
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Electrophysiology of Normal Cardiac Rhythm
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...