Related Experiment Video
Updated: Dec 20, 2025

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
PIN-PMN-PT Single Crystal 1-3 Composite-based 20 MHz Ultrasound Phased Array
Wei Zhou1,2,3, Tao Zhang1,3, Jun Ou-Yang1
1School of Optical and Electronic Information, Huazhong University of Science and Technology, Wuhan 430074, China.
Abstract:
Based on a modified dice-and-fill technique, a PIN-PMN-PT single crystal 1-3 composite with the kerf of 12 μm and pitch of 50 μm was prepared. The as-made piezoelectric composite material behaved with high piezoelectric constant (d33 = 1500 pC/N), high electromechanical coefficient (k = 0.81), and low acoustic impedance (16.2 Mrayls). Using lithography and flexible circuit method, a 48-element phased array was successfully fabricated from such a piezoelectric composite. The array element was measured to have a central frequency of 20 MHz and a fractional bandwidth of approximately 77% at -6 dB. Of particular significance was that this PIN-PMN-PT single crystal 1-3 composite-based phased array exhibits a superior insertion loss compared with PMN-PT single crystal and PZT-5H-based 20 MHz phased arrays. The focusing and steering capabilities of the obtained phased array were demonstrated theoretically and experimentally. These promising results indicate that the PIN-PMN-PT single crystal 1-3 composite-based high frequency phased array is a good candidate for ultrasound imaging applications.