Related Experiment Video
Updated: Dec 20, 2025

Intra-Operative Behavioral Tasks in Awake Humans Undergoing Deep Brain Stimulation Surgery
Published on: January 6, 2011
The Unsuccessful Twiddler: A Case of Twiddler's Syndrome Without Deep Brain Stimulator Lead Breakage
Hammad Ghanchi1, Taha M Taka2, Jacob E Bernstein1
1Neurosurgery, Riverside University Health System Medical Center, Moreno Valley, USA.
Abstract:
The authors present the case of a 78-year-old right-handed female with a past medical history of Parkinson's disease, treated with implantation of a left-sided subthalamic nucleus St. Jude Medical Infinity® (Abbott Medical, Austin, TX) deep brain stimulator, who presented with lead-associated discomfort, or "bowstringing". Further investigation by chest X-ray revealed an extensive case of distal lead coiling. However, it was surprising that, despite the extensive coiling, the lead stayed intact without hardware failure as proven by patient remaining asymptomatic from her Parkinson's disease and intraoperative impedance testing demonstrating normal results. After revision surgery, the patient remained asymptomatic. Due to paucity of cases of this disease in the literature, specific predictive risk factors are not known, but certain patient characteristics may help take precautions.