Related Experiment Video
Updated: Dec 20, 2025

A Rapid, Multiplex Dual Reporter IgG and IgM SARS-CoV-2 Neutralization Assay for a Multiplexed Bead-Based Flow Analysis System
Published on: April 6, 2021
The Allplex 2019-nCoV (Seegene) assay: which performances are for SARS-CoV-2 infection diagnosis?
Eric Farfour1, Philippe Lesprit2, Benoit Visseaux3
1Service de biologie Clinique, Hôpital Foch, Suresnes, France. e.farfour@hopital-foch.org.
Abstract:
Several commercial assays for SARS-CoV-2 RT-PCR are available but few of them were assessed. We evaluate the Allplex 2019-nCoV (Seegene) assay using 41 nasopharyngeal samples. The rates of agreement were 92.7% and 100% with the GeneFinder COVID-19 plus (Elitech) and the diagnosis of the infectious disease specialist respectively. Four samples display a Ct < 22.0 for the E and RdRp genes while the N gene was not detected, suggesting a variability of the viral sequence. There was no cross-reactivity with other respiratory viruses. The Allplex 2019-nCoV appears as a reliable method, but additional evaluations using more samples are needed. RT-PCR assays should probably include at least 2 viral targets.

