Determination of the Relative Cell Surface and Total Expression of Recombinant Ion Channels Using Flow Cytometry
Published on: September 28, 2016
Fluorometric Na+ Evaluation in Single Cells Using Flow Cytometry: Comparison with Flame Emission Assay.
Valentina E Yurinskaya1, Nikolay D Aksenov1, Alexey V Moshkov1
1Institute of Cytology, Russian Academy of Sciences, St. Petersburg, Russia.
Accurate intracellular sodium measurement using Asante Natrium Green-2 (ANG) requires careful calibration. ANG fluorescence is unreliable for quantifying intracellular sodium when ionophores are used for calibration, necessitating alternative methods for precise cellular sodium analysis.
Area of Science:
- Cellular Biology
- Biochemistry
- Analytical Chemistry
Background:
- Intracellular sodium concentration is crucial for fundamental cell functions.
- Fluorescent probes like Asante Natrium Green-2 (ANG) are vital for monitoring intracellular sodium in living cells.
- Current calibration methods for fluorescent sodium probes in situ using ionophores can lead to inaccurate results.
Purpose of the Study:
- To compare intracellular sodium concentration measurements using the ANG fluorescent probe and flame emission photometry.
- To investigate the impact of ionophores on ANG fluorescence and its correlation with actual intracellular sodium levels.
- To determine the reliability of ANG-2 for quantitative intracellular sodium assessment.
Main Methods:
- Intracellular sodium concentration in U937 cells was measured using flow cytometry with the ANG probe.
- Flame emission photometry combined with cellular water determination was used as a reference method.
- Intracellular sodium levels were manipulated using ionophores, ouabain (sodium pump inhibitor), or staurosporine (apoptosis inducer).
Main Results:
- Both ANG probe and flame photometry methods showed comparable results when sodium levels were altered by ouabain or staurosporine.
- ANG fluorescence was approximately two times lower in cells treated with ionophores compared to those with similar sodium concentrations but without ionophores.
- A significant discrepancy exists in ANG fluorescence when ionophores are used during calibration, impacting quantitative sodium assessment.
Conclusions:
- The sodium-sensitive dye ANG-2 is effective for detecting changes in intracellular sodium content and concentration in single cells and microparticles.
- ANG fluorescence, when calibrated in the presence of ionophores, does not accurately reflect the true intracellular sodium concentration.
- The observed reduction in ANG fluorescence in the presence of ionophores requires consideration for accurate quantitative intracellular sodium analysis.
More Related Videos
19:26Measuring Intracellular Ca2+ Changes in Human Sperm using Four Techniques: Conventional Fluorometry, Stopped Flow Fluorometry, Flow Cytometry and Single Cell Imaging
Published on: May 24, 2013
08:17Fluorescent Nanoparticles for the Measurement of Ion Concentration in Biological Systems
Published on: July 4, 2011
Related Concept Videos
Flame Photometry: Lab
Flame Photometry: Overview
Flow Cytometry
In...