Mirror symmetric broadband homodecoupled perfect echo spectroscopy

Peyman Sakhaii1, Bojan Bohorc1, Uwe Schliedermann1

  • 1NMR Laboratory of SANOFI, TIDES Analytical Sciences / German NMR Platform Industriepark Hoechst, Building G849, D-65926 Frankfurt/Main, Germany.

Summary

A new method enhances Total Correlation Spectroscopy (TOCSY) by extending the perfect echo filter, enabling longer acquisition times for improved homodecoupled spectra quality. This technique provides pure absorptive lineshapes and reduces proton spin multiplicities.

Related Concept Videos

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations01:08

IR Spectrum Peak Splitting: Symmetric vs Asymmetric Vibrations

Identical bonds within a polyatomic group can stretch symmetrically (in-phase) or asymmetrically (out-of-phase). Similar to hydrogen bonding, these vibrations also influence the shape of the IR peak. Generally, asymmetric stretching frequencies are higher than symmetric stretching frequencies. For example, primary amines exhibit two distinct IR peaks between 3300–3500 cm−1 corresponding to the symmetric and asymmetric N-H stretching, while secondary amines exhibit a single...
1.6K
2D NMR: Homonuclear Correlation Spectroscopy (COSY)01:06

2D NMR: Homonuclear Correlation Spectroscopy (COSY)

Homonuclear correlation spectroscopy, or COSY, is a 2-dimensional NMR technique that provides information about coupled protons. Typically, the geminal and vicinal coupling are observed. For example, consider the COSY spectrum of ethyl acetate, where its 1D proton NMR spectrum is plotted along the vertical and horizontal axes with their corresponding chemical shift scale. Three spots on the diagonal corresponding to the three peaks in the 1D proton spectrum are called diagonal peaks. The COSY...
1.8K
¹³C NMR: ¹H–¹³C Decoupling01:04

¹³C NMR: ¹H–¹³C Decoupling

The probability of having two carbon-13 atoms next to each other is negligible because of the low natural abundance of carbon-13. Consequently, peak splitting due to carbon-carbon spin-spin coupling is not observed in spectra. However, protons up to three sigma bonds away split the carbon signal according to the n+1 rule, resulting in complicated spectra.
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
1.6K
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)01:19

2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)

Heteronuclear single-quantum correlation spectroscopy (HSQC) is a 2D NMR technique that reveals one-bond correlations between hydrogen and a heteronucleus. The HSQC experiment is similar to the heteronuclear correlation experiment (HETCOR) but is more sensitive. In the HSQC spectrum, the proton chemical shift is plotted on the horizontal F2 axis, while the 13C chemical shift is plotted on the vertical F1 axis. The corresponding proton and 13C spectra are also shown. The HSQC contour plot does...
1.3K
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)01:20

¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)

When proton-coupled carbon-13 spectra are simplified by a broadband proton decoupling technique, structural information about the coupled protons is lost. Distortionless enhancement by polarization transfer (DEPT) is a technique that provides information on the number of hydrogens attached to each carbon in a molecule. While the DEPT experiment utilizes complex pulse sequences, the pulse delay and flip angle are specifically manipulated. The resulting signals have different phases depending on...
1.5K
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation01:26

Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation

Inductively coupled plasma (ICP) is the common plasma source used in atomic emission spectroscopy (AES), a technique that detects and analyzes various elements in a sample. This method is often called inductively coupled plasma atomic emission spectroscopy (ICP-AES).
There are three main types of inductively coupled plasma atomic emission spectroscopy  (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
533