Lack of functional integration and high manufacturing costs have been identified as major challenges in commercialization of point-of-care devices. In this study, roll-to-roll (R2R) fabrication process was developed for large-scale manufacturing of disposable microfluidic devices. The integrated, user-friendly device included a plasma separation membrane for simple blood filtration, immobilized antibodies for specific immunodetection, microfluidics for plasma transport and reagent mixing, and a blister for actuation and waste storage. These functionalities were designed to be compatible with R2R processing, which was demonstrated using pilot-scale printing lines producing 60 devices in an hour. The produced sensors enabled rapid (10 min) and sensitive (2 μg/mL) fluorescence-based immunodetection of C-reactive protein from 20 μL of whole blood.