Establishment and application of a new serum sodium candidate reference method
Xiaoyu Fan1, Qing Li1, Zhonggan Jin1
1Shanghai Center for Clinical Laboratory, Shanghai, PR China.
Summary
A new inductively coupled plasma mass spectrometry (ICP-MS) method accurately quantifies serum sodium. This reliable ICP-MS method offers a potential reference for standardizing sodium measurements in clinical labs.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Mass Spectrometry
Background:
- Serum sodium (Na) levels are critical clinical diagnostics.
- Accurate sodium quantification requires reliable laboratory methods.
- Evaluation of a novel inductively coupled plasma mass spectrometry (ICP-MS) method for serum sodium measurement.
Purpose of the Study:
- To assess the analytical performance of a new ICP-MS method for serum sodium quantification.
- To compare the ICP-MS method with existing sodium measurement techniques.
- To determine the suitability of the ICP-MS method as a reference method.
Main Methods:
- Serum samples were diluted and analyzed using ICP-MS with germanium (Ge) as an internal standard.
- Sodium calibration solutions were used with a standard addition method for quantification.
- Analytical performance, including precision and accuracy, was evaluated.
Main Results:
- The ICP-MS method demonstrated excellent linearity (r > 0.9999) and low coefficients of variation (<0.67%).
- Trueness was confirmed using a certified reference material (SRM 956d).
- Results showed high correlation (r = 0.9914) with the indirect ion-selective electrode (ISE) method, with a mean deviation of -0.15%.
Conclusions:
- The developed ICP-MS method is accurate, reliable, simple, and fast.
- This method is a strong candidate for a reference method in serum sodium standardization.
- It can aid in setting target values for routine laboratory measurements.
More Related Videos
Related Concept Videos
Clinically Relevant Drug Product Specifications: Methods of Establishment
119
Product specifications define the acceptable quality of a pharmaceutical product by ensuring identity, purity, potency, and strength. These specifications serve as benchmarks during development, manufacturing, and post-approval quality control. Clinically relevant specifications are particularly important because they directly relate to a drug's safety and efficacy in clinical use.Dissolution studies are critical biopharmaceutic tools that link in vitro behavior to in vivo performance. They...
119
Capillary Electrophoresis: Applications
927
Capillary electrophoretic separations offer various modes, each with unique applications. These modes include capillary zone electrophoresis, capillary gel electrophoresis, capillary array electrophoresis, capillary isoelectric focusing, capillary isotachophoresis, micellar electrokinetic chromatography, and capillary electrochromatography.
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
927