Validation Study to Determine the Accuracy of Central Blood Pressure Measurement Using the Sphygmocor Xcel Cuff
Martin G Schultz1, Dean S Picone1, Matthew K Armstrong1
1From the Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia (M.G.S., D.S.P., M.K.A., J.A.B., N.D., P.R.-T., J.E.S.).
Abstract:
Numerous devices purport to measure central (aortic) blood pressure (BP) as distinct from conventional brachial BP. This validation study aimed to determine the accuracy of the Sphygmocor Xcel cuff device (AtCor Medical, CardieX, Sydney, Australia) for measuring central BP. 296 patients (mean age 61±12 years) undergoing coronary angiography had simultaneous measurement of invasive central BP and noninvasive cuff-derived central BP using the Xcel cuff device (total n=558 individual comparisons). A subsample (n=151) also had invasive brachial BP measured. Methods were undertaken according to the Artery Society recommendations, and several calibration techniques to derive central systolic BP (SBP) were examined. Minimum acceptable error was ≤5±≤8 mm Hg. Central SBP was significantly underestimated, and with wide variability, when using the default calibration of brachial-cuff SBP and diastolic BP (DBP; mean difference±SD, -7.7±11.0 mm Hg). Similar variability was observed using other calibration methods (cuff 33% form-factor mean arterial pressure and DBP, -4.4±11.5 mm Hg; cuff 40% form-factor mean arterial pressure and DBP, 4.7±11.9 mm Hg; cuff oscillometric mean arterial pressure and DBP, -18.2±12.1 mm Hg). Only calibration with invasive central integrated mean arterial pressure and DBP was within minimal acceptable error (3.3±7.5 mm Hg). The difference between brachial-cuff SBP and invasive central SBP was 3.3±10.7 mm Hg. A subsample analysis to determine the accuracy of central-to-brachial SBP amplification showed this to be overestimated by the Xcel cuff device (mean difference 4.3±9.1 mm Hg, P=0.02). Irrespective of cuff calibration technique, the Sphygmocor Xcel cuff device does not meet the Artery Society accuracy criteria for noninvasive measurement of central BP.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
14:28Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
Published on: June 27, 2025
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Errors occurring during blood pressure monitoring
Several factors...