P- and R-wave Amplitude Sensed by Reveal LINQ™ Loop Recorder in Pediatric Patients
Roshan D'Souza1, Elizabeth Thomas2, Scott Macicek2
1Rainbow Babies and Children's Hospital, Case Western Reserve University School of Medicine, Cleveland, OH.
Abstract:
Implantable loop recorders are commonly used to sense arrhythmias. The purpose of this study is to assess the P- and R-wave amplitudes at implantation (I) and follow-up (F) following insertion of the Reveal LINQ™ Insertable Cardiac Monitor (Medtronic, Minneapolis, MN) in an institutional review board-approved, multicenter study performed on pediatric patients younger than 18 years old. Collected data included demographics, presence of congenital heart disease (CHD), P- and R-wave-sensed amplitude at I and F, and the method of implant (i.e. mapping or standard.) P waves were manually measured and R-wave sensing was recorded by the device. A total of 87 patients had a Reveal LINQ™ (Medtronic, Minneapolis, MN) device implanted; the mean patient age was 11.8 years (0.5 years to 18 years) with 48% of patients being female and 19% of patients having CHD; mapping was used in 43% of patients. The Reveal LINQ™ (Medtronic, Minneapolis, MN) experienced no change in average sensed R-wave amplitude at either I or F (1.28 mV vs 1.26 mV, p = NS). There was no difference in sensed R-wave amplitude noted with or without mapping used at I (1.29 mV vs 1.26 mV, p = NS) or F (1.48 mV vs 1.18 mV, p = NS). Additionally, no difference could be found in R-wave sensing of patients with CHD or without CHD at I (1.26 mV vs 1.4 mV, p = NS) or F (1.32 mV vs 1.32 mV, p = NS). R-wave sensing trended towards being inversely proportional to patient body surface area (BSA) (p = NS). P waves were detected on 48% of tracings in all patients at I and/or F, irrespective of whether the Reveal LINQ™ (Medtronic, Minneapolis, MN) device was placed with mapping. The R wave was (0.37-3.5 mV) at I and (0.3-3 mV) (p = NS) at F when P waves were detected. From these results, it can be said that the Reveal LINQ™ Insertable Cardiac Monitor (Medtronic, Minneapolis, MN) has an excellent ability to sense R-wave amplitude in pediatric patients. No significant difference in the sensing ability of the device could be identified with respect to the presence of CHD, use of mapping or BSA. P waves tended to be identified when there was a higher baseline R-wave amplitude.
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Special considerations while measuring pulse