Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Christoph Stotter1, Christoph Bauer2, Bojana Simlinger3
1Faculty of Health and Medicine, Department for Health Sciences, Medicine and Research, Center for Regenerative Medicine, Danube University Krems; Department of Orthopedics and Traumatology, LK Baden-Mödling-Hainburg; christoph.stotter@donau-uni.ac.at.
Abstract:
Osteochondral defects in middle-aged patients might be treated with focal metallic implants. First developed for defects in the knee joint, implants are now available for the shoulder, hip, ankle and the first metatarsalphalangeal joint. While providing pain reduction and clinical improvement, progressive degenerative changes of the opposing cartilage are observed in many patients. The mechanisms leading to this damage are not fully understood. This protocol describes a tribological experiment to simulate a metal-on-cartilage pairing and comprehensive analysis of the articular cartilage. Metal implant material is tested against bovine osteochondral cylinders as a model for human articular cartilage. By applying different loads and sliding speeds, physiological loading conditions can be imitated. To provide a comprehensive analysis of the effects on the articular cartilage, histology, metabolic activity and gene expression analysis are described in this protocol. The main advantage of tribological testing is that loading parameters can be adjusted freely to simulate in vivo conditions. Furthermore, different testing solutions might be used to investigate the influence of lubrication or pro-inflammatory agents. By using gene expression analysis for cartilage-specific genes and catabolic genes, early changes in the metabolism of articular chondrocytes in response to mechanical loading might be detected.