Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Comparison of Inflammatory Cell-Induced Corrosion and Electrocautery-Induced Damage of Total Knee Implants
Griffin Heise1, Christina Arnholt2, Jameson H Sorrels1
1Department of Orthopaedic Surgery and Biomedical Engineering, School of Medicine, University of Tennessee Health Science Center, Memphis TN.
Abstract:
Recently, inflammatory cell-induced corrosion (ICIC), a unique type of damage, has been reported in cobalt-chromium (CoCr) implants, but the mechanism remains poorly understood and controversial because electrocautery damage has also been shown to produce similar findings. This study aimed to distinguish between these two damage mechanisms. Forty-one CoCr primary total knee arthroplasty specimens were collected at time of necropsy. After removal and cleaning, light microscopy was used to identify areas of ICIC-like damage scars. A CoCr knee implant was intentionally damaged by electrocautery from both Bovie and Aquamantys sources using a 3-second hover method with 3 different energy settings for comparison to necropsy findings. Average roughness (Ra), max peak-to-valley height (Rmax), kurtosis (Rk), and skewness (Rsk) measurements were collected to represent the topography on the damaged areas for the CoCr implants. Necropsy implants showed signs of ICIC in 7 of 41 implants (17%) examined. Fe/C ratios of the Bovie electrocautery-damaged knee implant were shown to be statistically higher than those of necropsy-retrieved implants. Median Ra measurements were statistically less (P = 0.008) for Bovie-damaged areas compared to ICIC-dam-aged areas on CoCr. Median Rmax and Ra measurements were statistically less (P = 0.012, P < 0.001, respectively) for Aquamantys-damaged areas compared to ICIC-damaged areas on CoCr. While the visual patterns seen in necropsy-retrieved implants appeared similar to those with the intentionally damaged CoCrMo implant, the contents of the corroded regions are unique. The difference in roughness found on ICIC-damaged and electrocautery-damaged regions also indicates examination of surface topography as another distinguishing feature between the two mechanisms.