Related Experiment Video
Updated: Dec 20, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
External Fixation Devices Within the Magnetic Resonance Imaging Bore: A Safety and Radiologic Analysis
Scott Ryan1, Andrew S Moon1, Matthew Gordon1
1Department of Orthopaedic Surgery, Tufts Medical Center, Boston, MA.
Objectives:
To (1) report the thermal changes encountered at the pin/skin interface in a cadaver with a knee-spanning external fixator inside the magnetic resonance imaging (MRI) bore and (2) report on the quality of the MRI sequences collected.
Methods:
Three commonly used external fixation systems were placed on cadaveric lower extremities to simulate knee external fixation. Fiber optic thermal probes were placed at the pin/skin interface of a femoral and tibial pin. A control probe was embedded in the soft tissues of the thigh. Full knee MRI scans were performed using a 1.5-Tesla magnet. Real-time thermal data were collected. A clinically significant increase in temperature compared with the control was defined as 2°C. Two blinded radiologists evaluated the images for image quality and overall diagnostic utility using a standardized 5-point grading scale.
Results:
There were statistically significant differences in the temperature changes between the femoral/tibial pin sites and the control probe sites during each phase of the MRI scan. However, there was only one clinically significant difference in temperature change during a single sequence of one MRI scan of one of the external fixator devices. Overall image quality was graded as a 4 for each image set with 100% interobserver agreement (k = 1.0).
Conclusions:
Despite significant differences in temperature changes between the pin sites and controls over multiple MRI sequences in commonly used external fixator devices, the differences in temperature change are likely not clinically relevant. Overall image quality and interpretability of the images were excellent.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...