Related Experiment Video
Updated: Dec 19, 2025

Author Spotlight: Development of a Smartphone-Enhanced Paper-Based Device for Rapid Dengue NS1 Detection
Published on: January 26, 2024
First successful field evaluation of new, one-minute haemozoin-based malaria diagnostic device
Rajat Kumar1, Anil K Verma1, Shweta Shrivas1
1ICMR-National Institute of Research in Tribal Health (NIRTH), Jabalpur, Madhya Pradesh, India.
Background:
Early and accurate diagnosis of malaria is critical to the success of malaria elimination. However, the current mainstay of malaria diagnosis in the field, such as light microscopy and rapid diagnostic tests (RDTs), have limitations due to low parasite density or mutation in diagnostic markers.
Methods:
We evaluated an inexpensive, robust, rapid, malaria diagnostic device, called Gazelle, that employs magneto-optical detection to identify haemozoin crystals (Hz) produced by all species of human malaria parasites in infected individuals. A beam of polarised light is passed through the lysed diluted blood sample under the influence of high (~.55T) and low magnetic fields. The difference in light transmission through the sample between the high and low magnetic fields indicates presence of Hz, suggesting possible malarial infection. A total of 300 febrile patients were screened at the malaria clinic of Indian Council of Medical Research-National Institute of Research in Tribal Health (ICMR-NIRTH), Jabalpur, India, from August 2018 to November 2018. Malaria diagnosis was done using four diagnostic methods: Gazelle, light microscopy, RDT, and malaria specific Polymerase Chain Reaction (PCR). Measures of diagnostic accuracy were compared.
Findings:
Out of 300 febrile patients enroled and tested for the presence of malaria parasites, 262 patient samples were included in the final analysis. The sensitivity and specificity of Gazelle was 98% and 97% in comparison to light microscopy, 82% and 99% to PCR and 78% and 99% to RDT, respectively. The results of the four diagnostic methods were comparable and statistically no significant differences in sensitivity or specificity was observed between these methods. Enhanced diagnostic accuracy of Gazelle in malaria patients with no prior history of malaria treatment was observed in this study.
Interpretation:
The diagnostic ability of Gazelle was comparable to light microscopy and better than RDTs even in low parasitemia and in presence of pfhrp2/3 deletion mutant parasites. Gazelle may be a novel valuable diagnostic tool in resource poor settings where (i) microscopy is not feasible and (ii) pfhrp2/3gene deleted parasite are present. Its speed, cost-efficiency, and alternative to lack of microscopists makes it an important adjunct in field settings.
Funding:
HemexDx, India.
More Related Videos
09:00A Multi-detection Assay for Malaria Transmitting Mosquitoes
Published on: February 28, 2015
10:50Detection and Quantification of Plasmodium falciparum in Aqueous Red Blood Cells by Attenuated Total Reflection Infrared Spectroscopy and Multivariate Data Analysis
Published on: November 2, 2018