Background:
A linear relationship between baseplate insertion torque and compression force in reverse shoulder arthroplasty (RSA) baseplates with central screw design has been recently established. In this study, we evaluated 3 different baseplate designs and their influence on the torque-compression relationship.
Methods:
Three different RSA baseplate designs were evaluated through biomechanical testing using a glenoid vault, bone surrogate model. A digital torque gauge was used to measure insertion torque applied to the baseplate, whereas compression data were collected continuously from a load cell. Additionally, 2 predictive models were developed to predict the compression forces of each baseplate design at varying levels of torque.
Results:
A linear relationship was found between baseplate compression and insertion torque for all 3 baseplate designs. Both the monoblock and 2-piece locking designs achieved the goal torque of 6.8 Nm, whereas the 2-piece nonlocking design did not due to material strip-out. No significant difference in maximum compression was found between the monoblock and 2-piece locking designs. However, the 2-piece nonlocking design achieved significantly higher compression. Both predictive models were shown to adequately predict compressive forces at different torque inputs for the monoblock and 2-piece locking designs but not the 2-piece nonlocking design.
Conclusion:
The torque-compression relationship of a central screw baseplate is significantly affected by baseplate design. A 2-piece nonlocking baseplate reaches higher compression levels and risks material strip-out at lower insertional torques compared with a monoblock and 2-piece locking design. This has implications both on component design and on surgeon tactile feedback during surgery.
Related Concept Videos
Residual Stresses in Circular Shafts
Screw: Problem Solving
To evaluate the...
Design of Transmission Shafts - Stress Analysis
Circular Shafts - Elastoplastic Materials
As torque on the...
Torque
Torque can be considered as the rotational counterpart to force. Since forces change the translational...
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...