Radiation dose and image quality of CT fluoroscopy with partial exposure mode
Keisuke Takiguchi1, Atsushi Urikura2, Tsukasa Yoshida1
1Radiation and Proton Therapy Center, Shizuoka Cancer Center, Shizuoka, Japan.
Purpose:
The present study aimed to evaluate the scan technique of computed tomography (CT)-guided puncture procedures using partial exposure mode (PEM) on the radiation dose of the operator's hand and image quality.
Methods:
Radiation dose was evaluated using three types of scanning methods: one-shot scan (OS), OS with a bismuth shield added (OSBismuth), and a half-scan (i.e., PEM) capable of an adjustable exposure angle. Dose evaluation was performed using a torso phantom, while a circular phantom simulating the liver parenchyma and lesions was used for image quality evaluation. For each scanning method, four measurements were made to determine the radiation dose to the operator's hand and the dose distribution on the surface of the patient's torso; the output-dose profile was determined from five measurements. Image quality was evaluated in terms of contrast and contrast-to-noise ratio (CNR). Analysis of variance (ANOVA) or Friedman test were used for comparison between groups as appropriate. The post hoc tests were Tukey's honestly difference (HSD) test for parametric data or Wilcoxon signed rank test with Bonferroni correction for nonparametric data.
Results:
The PEM yielded a radiation dose to the operator's hand that was 84% (0.35 vs. 2.33 mGy) lower than that of the OS. The dose to the patient's torso was reduced by 35% and 68% for the OSBismuth and PEM, respectively, relative to that of the OS. Compared with the CNR of the other two scanning methods (OS, 2.9±0.1; OSBismuth, 2.9±0.1), the PEM increased the standard deviation and decreased the CNR (2.1±0.04, Tukey's HSD, P < 0.001 for all). Images acquired with PEM showed visibility equivalent to that of other scanning methods when window conditions were adjusted.
Conclusion:
This study demonstrated that CT-guided puncture procedure using PEM effectively reduces the operator's exposure to radiation while minimizing image quality deterioration.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT