Accuracy of the Hand-held Wavefront Aberrometer in Measurement of Refractive Error
Jae Yong Han1, Sangchul Yoon1,2, Nicolas Scott Brown3
1Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.
Purpose:
To compare refractive error measured by hand-held wavefront aberrometers with postcycloplegic autorefraction (AR) and cycloplegic refraction (CR).
Methods:
The medical records of patients who received refractive measurements using the wavefront aberrometer, postcycloplegic AR, and CR between January 2014 and January 2016 were retrospectively analyzed. The mean differences, 95% confidence intervals, and limits of agreement (LOA) were calculated for the refractive vector components (M, J₀, and J ).
Results:
Fifty-one patients (9.0 ± 5.5 years, male 41.2%) were enrolled in this study, and only the right eye of each was included. Refractive errors ranged from -9.25 to +7.25 diopters (D) for spherical equivalent (median, 0.75 D). The M component was not significantly different among the three methods (p = 0.080). However, the J₀ vector component was significantly different (p < 0.001). After post hoc analysis, the wavefront aberrometer obtained more positive values for J₀ compared to the other methods. The J component was not significantly different among the three methods (p = 0.143). The mean difference between the wavefront aberrometer and postcycloplegic AR was -0.115 D (LOA, -1.578 to 1.348 D) for M, 0.239 D (LOA, -0.371 to 0.850 D) for J₀, and -0.015 D (LOA, -0.768 to 0.738 D) for J . The mean difference between the wavefront aberrometer and CR was -0.220 D (LOA, -1.790 to 1.350 D) for M, 0.300 D (LOA, -0.526 to 1.127 D) for J₀, and -0.079 D (-0.662 to 0.504 D) for J .
Conclusions:
The wavefront aberrometer showed good agreement with postcycloplegic AR and CR in spherical equivalents, but tended to produce slightly myopic results. The wavefront aberrometer also overestimated with-the-rule astigmatism. Therefore, we recommend that the device be used for estimations of refractive error, which may be useful for patients who have postural difficulties, live in undeveloped countries, or are bedridden.