Evaluation of VIASURE Sexually Transmitted Diseases Real Time PCR Detection kit (CerTest Biotec) for the diagnostic

Cristina Casañ López1, Belén Rivaya Sánchez1, Gema Fernández Rivas2

  • 1Servicio de Microbiología, Laboratorio Clínico Metropolitana Norte, Hospital Universitari Germans Trias i Pujol, Badalona, Spain.

Abstract

Insights

The VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit shows high agreement with the reference assay for diagnosing STIs. This real-time PCR test is a reliable tool for clinical laboratories, though results near the limit of detection require careful interpretation.

Area of Science:

  • Clinical Microbiology
  • Molecular Diagnostics
  • Infectious Diseases

Background:

  • Sexually transmitted infections (STIs) are a growing public health concern with increasing incidence.
  • Traditional diagnostic methods for STIs have limitations; real-time PCR offers improved accuracy and speed.
  • Real-time PCR has become a cornerstone in modern microbiology laboratories for sensitive and specific pathogen detection.

Purpose of the Study:

  • To evaluate the performance of the VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit.
  • To compare the VIASURE kit against an established reference real-time PCR assay (Allplex™ STI7 Essential Assay).
  • To assess the utility of the VIASURE kit as a diagnostic tool for STIs.

Main Methods:

  • Analysis of 948 clinical samples (vaginal, endocervical, urethral, rectal, pharyngeal swabs, and urine) collected between July and September 2018.
  • Comparative evaluation using the VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit and the Allplex™ STI7 Essential Assay as the reference method.
  • Assessment of discordant results and agreement using kappa (k) coefficient analysis.

Main Results:

  • A high overall agreement was observed between the VIASURE kit and the reference assay, with kappa coefficients ranging from 0.92 to 0.99.
  • Discordant results were found in 4.5% (43/948) of samples, primarily associated with threshold cycle (Ct) values near the limit of detection.
  • The VIASURE kit demonstrated robust performance across various sample types.

Conclusions:

  • The VIASURE Sexually Transmitted Diseases Real-Time PCR Detection kit exhibits excellent correlation with the Allplex STI7 assay, confirming its suitability for STI diagnostics.
  • The kit is a valuable tool for clinical laboratories managing the increasing burden of STIs.
  • Caution is advised when interpreting positive results with high Ct values (≥35) or low amplification signals, necessitating correlation with clinical data.