Related Experiment Video
Updated: Dec 19, 2025

12:03
A Method for Tracking the Time Evolution of Steady-State Evoked Potentials
Published on: May 25, 2019
8.8K
His bundle pacing using a simple stylet and a standard active fixation electrode
Serkan Cay1, Aysenur Ekizler1, Emin Karimli1
1Division of Arrhythmia and Electrophysiology, Department of Cardiology, University of Health Sciences, Ankara City Hospital, Ankara, Turkey.
Journal of Electrocardiology
|June 7, 2020
Abstract:
Conventionally, His bundle pacing (HBP) is achieved using specially designed pacing leads and delivery sheaths. This paper describes the feasibility of permanent HBP with a pre-shaped simple stylet and a standard active-fixation electrode, through axillary vein access, without using dedicated delivery tools. This method may be a feasible and safe alternative to the only commercially available system.