Related Experiment Video
Updated: Dec 18, 2025

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Patient-specific instrumentation does not improve tibial component coronal alignment for medial UKA compared to
Houssam Kalache1, Jacobus H Müller2, Mo Saffarini3
1Centre Hospitalier Privé Saint-Grégoire, 6 Boulevard de la Boutière, 35760, Saint-Grégoire, France.
Background:
Patient-specific instrumentation (PSI) may potentially improve unicompartmental knee arthroplasty (UKA) implant positioning and alignment. The purpose of this study was to compare early radiographic coronal alignment of medial UKA performed using PSI versus conventional instrumentation (CI) for tibial resections.
Methods:
A consecutive series of 47 knees (47 patients) received medial UKA, with the tibial resections performed using CI (first 22 knees) or PSI (next 25 knees), while femoral resections were performed with CI in both groups. The target mechanical medial proximal tibial angle (mMPTA) was 87° ± 3°, and the target hip-knee-ankle (HKA) angle was 177° ± 2°. The postoperative mMPTA and HKA were evaluated from postoperative radiographs at a follow-up of 2 months.
Results:
Differences in postoperative mMPTA (p = 0.509) and HKA (p = 0.298) between the two groups were not statistically significant. For the mMPTA target, 24% of knees in the PSI group (85.6° ± 2.1°) and 32% of the CI group (85.0° ± 3.6°) were outliers. For the HKA target, 44% of knees in the PSI group (176.3° ± 2.8°) and 18% of the CI group (177.1° ± 2.3°) were outliers. Considering the two criteria simultaneously, 60% of knees in the PSI group and 45% of knees in the CI group were outside the target zone (p = 0.324), whereas 28% of knees in the PSI group and 41% of knees in the CI group were outside the target zone by more than 1° (p = 0.357).
Conclusions:
The results of the present study revealed no statistically significant difference in radiographic coronal alignment of UKA performed using PSI versus CI for tibial resections.

