A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Effect of Nitric Oxide Delivery Device on Tidal Volume Accuracy During Mechanical Ventilation at Small Tidal Volumes
Courtney D Ranallo1, Tracy L Thurman2, Shirley J Holt3
1Department of Pediatrics/Critical Care Medicine, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma. courtney-ranallo@ouhsc.edu.
Background:
Inhaled nitric oxide (INO) is used in infants as a therapy for elevated pulmonary vascular resistance. When INO is delivered at low tidal volumes, displayed inspiratory and expiratory volumes vary widely. We hypothesize that volume is removed by the sampling line during the ventilation cycle, and this results in a net volume loss at low tidal volumes. This study aimed to measure the volumes delivered and to assess the accuracy of displayed ventilator values using a test lung.
Methods:
A test lung was connected to a ventilator and an INO delivery system. All tests were performed with stable mode settings across volumes of 18, 30, 42, and 60 mL. Flow measured with a pneumotachometer attached between the test lung and the circuit assessed the percent error between inspiratory and expiratory volumes measured by the pneumotachometer measured and displayed on the ventilator under various INO/sample line conditions to determine where and how much volume was being displaced.
Results:
Displayed and measured inspiratory volumes had small variations between the INO/sample line conditions and baseline. However, expiratory volumes, with the sample line connected, exhibited large percent error values that increased (-14, -20, -27, and -34) as tidal volume decreased (60, 42, 30, and 18 mL) and error was significantly larger compared to baseline in all tidal volumes (P < .01) with and without INO delivery.
Conclusions:
We concluded that inspiratory volumes were not affected by INO delivery, but additional removal of volume in the expiratory phase of the breath cycle by the sampling line results in a large error in the displayed expiratory volume.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...