Related Experiment Video
Updated: Dec 18, 2025

The Supraclavicular Fossa Ultrasound View for Central Venous Catheter Placement and Catheter Change Over Guidewire
Published on: December 23, 2014
A calcified central venous fibrin sheath mimicking a fractured catheter
Debajyoti Roy1, Yeon Wenxiang1, Shrikant D Pande2
1Department of Renal Medicine, Changi General Hospital, Singapore, Singapore.
Abstract:
A 75-year-old female with end stage kidney failure had her tunneled central venous dialysis catheter (CVC) removed. A subsequent computed tomopgraphy (CT) scan of the chest reported a filling defect in the central vein that appeared to represent a fractured remnant of the CVC. The catheter had been retained for culture and was available for direct visualization, which showed it to be entirely intact. A subsequent venogram confirmed that the CT findings represented a retained calcified central venous fibrin sheath. As retained CVC fragments may require intervention, this diagnosis should be established carefully. A calcified fibrin sheath associated with a chronic CVC is a known, although rare, complication and should be considered in the differential diagnosis of an apparent CVC fracture prior to further interventions.
Related Concept Videos
Venous Thrombosis I: Introduction
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis III: Interprofessional Care
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview