Source-Detector Spectral Pairing-Related Inaccuracies in Pulse Oximetry: Evaluation of the Wavelength Shift
Olivier Tsiakaka1,2, Benoit Gosselin1,2, Sylvain Feruglio3
1Department of Computer and Electrical Engineering, Université Laval, 1065 Avenue de la Médecine, Quebec, QC G1V 0A6, Canada.
Abstract:
Pulse oximetry enables oxygen saturation estimation ( S p O 2) non-invasively in real time with few components and modest processing power. With the advent of affordable development kits dedicated to the monitoring of biosignals, capabilities once reserved to hospitals and high-end research laboratories are becoming accessible for rapid prototyping. While one may think that medical-grade equipment differs greatly in quality, surprisingly, we found that the performance requirements are not widely different from available consumer-grade components, especially regarding the photodetection module in pulse oximetry. This study investigates how the use of candidate light sources and photodetectors for the development of a custom S p O 2 monitoring system can lead to inaccuracies when using the standard computational model for oxygen saturation without calibration. Following the optical characterization of selected light sources, we compare the extracted parameters to the key features in their respective datasheet. We then quantify the wavelength shift caused by spectral pairing of light sources in association with photodetectors. Finally, using the widely used approximation, we report the resulting absolute error in S p O 2 estimation and show that it can lead up to 8% of the critical 90-100% saturation window.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...