Related Experiment Video
Updated: Dec 18, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Prospective Assessment of an Automated Intraprocedural 12-Lead ECG-Based System for Localization of Early Left
Shijie Zhou1,2,3, Amir AbdelWahab3, B Milan Horáček4
1Department of Biomedical Engineering (S.Z., N.A.T.), Johns Hopkins University, Baltimore, MD.
Background:
To facilitate ablation of ventricular tachycardia (VT), an automated localization system to identify the site of origin of left ventricular activation in real time using the 12-lead ECG was developed. The objective of this study was to prospectively assess its accuracy.
Methods:
The automated site of origin localization system consists of 3 steps: (1) localization of ventricular segment based on population templates, (2) population-based localization within a segment, and (3) patient-specific site localization. Localization error was assessed by the distance between the known reference site and the estimated site.
Results:
In 19 patients undergoing 21 catheter ablation procedures of scar-related VT, site of origin localization accuracy was estimated using 552 left ventricular endocardial pacing sites pooled together and 25 VT-exit sites identified by contact mapping. For the 25 VT-exit sites, localization error of the population-based localization steps was within 10 mm. Patient-specific site localization achieved accuracy of within 3.5 mm after including up to 11 pacing (training) sites. Using 3 remotes (67.8±17.0 mm from the reference VT-exit site), and then 5 close pacing sites, resulted in localization error of 7.2±4.1 mm for the 25 identified VT-exit sites. In 2 emulated clinical procedure with 2 induced VTs, the site of origin localization system achieved accuracy within 4 mm.
Conclusions:
In this prospective validation study, the automated localization system achieved estimated accuracy within 10 mm and could thus provide clinical utility.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
08:19Electrophysiological Assessment of Murine Atria with High-Resolution Optical Mapping
Published on: February 22, 2018
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...