1024-pixel image matrix for chest CT - Impact on image quality of bronchial structures in phantoms and patients
André Euler1, Katharina Martini1,2, Bettina Baessler1
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Objectives:
To compare objective and subjective image quality of bronchial structures between a 512-pixel and a 1024-pixel image matrix for chest CT in phantoms and in patients.
Materials And Methods:
First, a two-size chest phantom was imaged at two radiation doses on a 192-slice CT scanner. Datasets were reconstructed with 512-, 768-, and 1024-pixel image matrices and a sharp reconstruction kernel (Bl64). Image sharpness and normalized noise power spectrum (nNPS) were quantified. Second, chest CT images of 100 patients were reconstructed with 512- and 1024-pixel matrices and two blinded readers independently assessed objective and subjective image quality. In each patient dataset, the highest number of visible bronchi was counted for each lobe of the right lung. A linear mixed effects model was applied in the phantom study and a Welch's t-test in the patient study.
Results:
Objective image sharpness and image noise increased with increasing matrix size and were highest for the 1024-matrix in phantoms and patients (all, P<0.001). nNPS was comparable among the three matrices. Objective image noise was on average 16% higher for the 1024-matrix compared to the 512-matrix in patients (P<0.0001). Subjective evaluation in patients yielded improved sharpness but increased image noise for the 1024- compared to the 512-matrix (both, P<0.001). There was no significant difference between highest-order visible bronchi (P>0.07) and the overall bronchial image quality between the two matrices (P>0.22).
Conclusion:
Our study demonstrated superior image sharpness and higher image noise for a 1024- compared to a 512-pixel matrix, while there was no significant difference in the depiction and subjective image quality of bronchial structures for chest CT.
More Related Videos
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT