Autopsy evaluation of the implantation site of a His bundle pacing lead demonstrating selective capture

Hiroyuki Kato1, Osamu Igawa2, Kazumasa Suga3

  • 1Division of Cardiology, Japan Community Health Care Organization Chukyo Hospital, Nagoya, Japan.

Insights

Autopsy revealed His bundle pacing lead placement through the tricuspid valve, causing a prolonged interval. Transthyretin amyloidosis may impact pacing effectiveness in elderly patients.

Area of Science:

  • Cardiovascular Medicine
  • Electrophysiology
  • Cardiac Pathology

Background:

  • His bundle pacing (HBP) is an advanced technique for heart rhythm management.
  • Autopsy evaluations of HBP lead placement are infrequent, limiting understanding of long-term lead behavior.
  • Atrioventricular block and heart failure are common indications for pacing in elderly individuals.

Observation:

  • A 98-year-old male patient underwent HBP implantation for atrioventricular block and heart failure.
  • Post-implantation, selective HBP was achieved, but with a prolonged stimulus-to-QRS interval.
  • Autopsy revealed the HBP lead positioned ventrally, traversing the anteroseptal commissure of the tricuspid valve.

Findings:

  • Macroscopic examination confirmed unusual lead placement, potentially explaining the extended stimulus-to-QRS duration.
  • The lead's passage through the tricuspid valve commissure may have affected conduction.
  • The patient's transthyretin cardiac amyloidosis could have contributed to distal conduction system abnormalities.

Implications:

  • This case highlights the importance of detailed autopsy in understanding HBP lead positioning and its impact on pacing parameters.
  • Transthyretin cardiac amyloidosis presents a potential challenge for achieving optimal His bundle pacing outcomes.
  • Further research is needed to correlate lead placement variations with clinical outcomes and cardiac pathology in HBP patients.