Glucose meters: What's the laboratory reference glucose?
Mariarosa Carta1, Davide Giavarina1, Andreina Paternoster1
1St. Bortolo Hospital, Clinical Laboratory, Vicenza, Italy.
Journal of Medical Biochemistry
|June 19, 2020
Summary
Accurate glucose meter evaluation requires laboratory reference glucose analysis. Using citrate-containing tubes ensures reliable results, unlike those with only NaF, which can be affected by in vitro glycolysis.
Area of Science:
- Clinical Chemistry
- Biomedical Engineering
- Diabetes Diagnostics
Background:
- Glucose meter accuracy is critical for diabetes management and is typically assessed against laboratory reference analyzers.
- Scientific societies recommend glycolysis inhibitors like citrate for precise glucose measurements.
- In vitro glycolysis can occur in blood samples, potentially affecting glucose concentration readings.
Purpose of the Study:
- To evaluate the bias between capillary and plasma glucose measurements using different anticoagulant tubes.
- To assess the accuracy of glucose meters against laboratory reference methods under varying conditions.
- To determine the impact of NaF and citrate mixtures in blood collection tubes on glucose determination.
Main Methods:
- 139 volunteers undergoing oral glucose tolerance tests (OGTT) were enrolled.
- Capillary glucose was measured using two different glucose meters (Abbott FreeStyle Precision Neo, Roche Cobas Accu-Chek Inform II).
- Plasma glucose was analyzed using a hexokinase method on Dimension Vista systems, with samples collected in tubes containing NaF/citrate mixtures (liquid and granular) or NaF alone.
Main Results:
- When using liquid and granular citrate mixture tubes, 96.82% and 97.37% of capillary glucose determinations met specifications, respectively.
- In contrast, only 73.02% and 80.26% of determinations were within specifications when using tubes containing only NaF.
- These findings highlight a significant difference in accuracy based on the anticoagulant used.
Conclusions:
- Accurate evaluation of glucose meter accuracy necessitates understanding the laboratory reference glucose method.
- The use of blood collection tubes containing only NaF can lead to erroneous assessments of glucose meter accuracy due to in vitro glycolysis.
- Tubes incorporating citrate mixtures are essential for the correct evaluation of glucose meter accuracy, ensuring reliable diagnostic outcomes.
Related Concept Videos
Hypoglycemia and Glucagon
698
Without prolonged fasting, healthy individuals maintain blood glucose levels above 3.5 mM due to a well-adapted neuroendocrine counterregulatory system that effectively prevents acute hypoglycemia, a potentially life-threatening condition. The primary clinical scenarios for hypoglycemia encompass diabetes treatment, inappropriate production of endogenous insulin or insulin-like substances by tumors, and the use of glucose-lowering agents in non-diabetic individuals. Notably, hypoglycemia in the...
698
Glucose Homeostasis: Regulation of Blood Glucose
3.5K
Carbohydrates consumed through foods are converted into glucose, a crucial energy source for the body. In the prandial state, high blood glucose levels stimulate the secretion of insulin from the pancreas. Insulin inhibits hepatic glucose production and stimulates glucose uptake and metabolism by muscle and adipose tissue. The excess glucose is converted into glycogen and stored in the liver and muscles.
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
3.5K
Diabetes Mellitus: Type 2 and Gestational
4.2K
Type 2 diabetes, characterized by insulin resistance, arises when the insulin receptors on cells lose responsiveness to insulin, diminishing the cell's capacity to take up glucose, resulting in elevated blood glucose levels. To receive a diagnosis of Type 2 diabetes, a series of blood glucose tests are necessary to assess whether the blood glucose falls within normal parameters. If the result is out of the normal range, a patient may be diagnosed as prediabetic or diabetic, depending on the...
4.2K
Glucose Transporters
27.0K
Glucose transporters facilitate the transport of glucose across the cell membrane. In addition to glucose, some glucose transporters can also aid the movement of other hexoses such as fructose, mannose, and galactose.
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
27.0K
Diabetes: Symptoms, Diagnosis, and Complications
1.9K
For most patients, experiencing several weeks of polyuria, polydipsia, fatigue, and significant weight loss may indicate the presence of diabetes. Furthermore, adults displaying the phenotypic appearance of type 2 diabetes (particularly those who are obese and not initially insulin-requiring), may have islet cell autoantibodies, suggesting autoimmune-mediated β cell destruction and a diagnosis of latent autoimmune diabetes of adults (LADA). The categorization of glucose homeostasis is...
1.9K