Are ultrasound foetal circumference measurement methods interchangeable?
1Radiation Protection & Radiology Physics, Lincoln County Hospital, Lincoln, UK.
Ultrasound (Leeds, England)
|June 19, 2020
Summary
Ultrasound measurements of fetal circumferences using ellipse fitting and derived methods are not interchangeable. The derived method is recommended for consistency with existing fetal size and weight charts.
Area of Science:
- Obstetrics and Gynecology
- Medical Imaging
- Fetal Biometry
Background:
- Ultrasound is crucial for fetal biometry, with ellipse fitting and derived measurements commonly used for circumferences.
- These methods are often assumed to be equivalent, despite potential differences impacting clinical practice.
- Accurate fetal measurements are vital for monitoring growth and estimating fetal weight.
Purpose of the Study:
- To investigate and quantify the differences between ellipse fitting and derived fetal circumference measurements in clinical settings.
- To assess the clinical implications of these differences on estimated fetal weight calculations.
- To provide guidance on appropriate method selection based on existing charting standards.
Main Methods:
- Retrospective analysis of fetal head and abdominal circumferences measured via ellipse fitting.
- Derivation of circumferences using a standard formula from anterior-posterior and transverse diameters.
- Comparison of ellipse fitted and derived measurements using Bland-Altman plots, stratified by gestational age groups.
Main Results:
- Small systematic differences were observed between ellipse fitted and derived measurements.
- Random differences varied significantly across gestational age groups, with wider confidence intervals in later scans.
- 95% confidence intervals for head circumference ranged from ±3% to ±6%, abdominal circumference from ±5% to ±6%, and estimated fetal weight from ±5% to ±12%.
Conclusions:
- Ellipse fitting and derived fetal circumference measurement methods are not interchangeable.
- The choice of method significantly impacts measurement accuracy and estimated fetal weight.
- The derived method should be preferentially used when referencing charts based on this methodology for fetal size, growth, and weight estimation.
Related Concept Videos
Ultrasonography
7.1K
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called...
During an ultrasonography procedure, a handheld device called...
7.1K
Ultrasound II: Endoscopic Ultrasound and FibroScan
398
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
398