Short-time evaluation on intraocular scattering after implantable collamer lens implantation for correcting high
1Tianjin Eye Hospital, Tianjin Key Lab of Ophthalmology and Visual Science, Tianjin Eye Institute, Clinical College of Ophthalmology, Tianjin Medical University, Nankai University Eye Hospital, No. 4 Gansu Road, Heping District, Tianjin, People's Republic of China.
Background:
To compare the intraocular scattering before and after implantation of implantable collamer lens (ICL) V4c for correction of high myopia in a short term.
Methods:
In this study, 38 eyes of 19 patients who underwent the implantation of ICL V4c were followed up for 3 months. Uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), objective scattering index (OSI), modulation transfer function cutoff frequency (MTF cutoff), strehl ratio (S/R), OV100%, OV20% and OV9% were measured pre- and postoperatively. Meantime, the Pseudophakic Dysphotopsia Questionnaire (PDQ) was scored to evaluate the subjective satisfaction of intraocular scattering pre- and postoperatively.
Results:
The UCVA were - 0.02 ± 0.06, - 0.03 ± 0.07 and - 0.04 ± 0.07 logMAR at 1 week, 1 month and 3 months postoperatively which were significantly better than those preoperatively (P < 0.05). The BCVA were - 0.09 ± 0.09, - 0.09 ± 0.1 and - 0.1 ± 0.11 logMAR at 1 week, 1 month and 3 months after surgery, which were better than those before surgery significantly (t = 15.64, P < 0.05). The mean OSI were 2.37 ± 1.6, 1.63 ± 0.94, 1.5 ± 0.86 and 1.43 ± 1.05 preoperatively, 1 week, 1 month and 3 months postoperatively which was found significant difference (F = 12.92 P < 0.05). No significant differences were found in MTF cut off (F = 0.61, P = 0.62), S/R (F = 0.58, P = 0.36), OV100% (F = 0.966, P = 0.65), OV20% (F = 0.121, P = 0.96) and OV9% (F = 1.01, P = 0.30) between pre- and postoperatively. The PDQ results indicated that intraocular scattering reduced at 3 months after surgery significantly (P < 0.05).
Conclusions:
The ICL V4c implantation for correcting high myopia induced less intraocular scattering and visual disorder than spectacle correction.