Characterizing a PTW microDiamond detector in kilovoltage radiation beams
Ahtesham Ullah Khan1, Wesley S Culberson1, Larry A DeWerd1
1Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, 53705, USA.
Purpose:
The aim of this work was to characterize the dosimetric properties of the PTW microDiamond (60019) single crystal synthetic diamond detector (DD) in kilovoltage x-ray beams. The following characteristics were addressed in this study: required preirradiation dose, dose-rate linearity, energy dependence, and percent depth dose response of the DD.
Methods:
UWADCL x-ray beams, characterized by NIST-traceable ionization chambers, were used in this study. Preirradiation dose required by the DD, in order to stabilize the detector's response to within 0.1%, was quantitated. Dose-rate dependence was also investigated using the UW250-M and UW50-M beams, where the dose rate was varied by changing the tube current. and calibration coefficients for all the available M series beams at UWADCL were obtained to determine the energy dependence of the DD, Diode E, Diode P, and P11 parallel-plate ionization chamber. A custom-built water tank was utilized to measure the percent depth dose (PDD) response of the DD, Diode E, Diode P, and P11 chamber in UW250-M, UW100-M, and UW50-M beams. The measured PDD response of the detectors was compared with the simulated PDD data using EGSnrc Monte Carlo code.
Results:
A 1.5 Gy dose-to-water or air-kerma was found to be sufficient for the given DD's response to stabilize to within 0.1% in all of the beams used in this study. The dose-rate dependence parameter, Δ, was found to be 1.00 ± 0.02 and 1.016 ± 0.05 for the UW250-M and UW50-M beams, respectively. Relative to the Co calibration coefficients, the DD was found to under-respond relative to calculated absorbed dose to water response and over-respond relative to the calculated air-kerma response in the M-series beams. Agreement of 1.5% was found between the measured PDD values and Monte Carlo simulated PDD values for UW250-M, UW100-M, and UW50-M beams.
Conclusions:
In order to stabilize the response, the DD needs a preirradiation dose, which is unique to every DD. A linear relationship between detector response and dose rate was found within the evaluated uncertainty. An energy dependence of the DD was studied, which is more pronounced in the low-energy beams and can be partially attributed to the metal contact material around the sensitive volume of the DD. Overall, the DD was found to be suitable for kilovoltage x-ray dosimetry.
More Related Videos
06:51Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...