Objectives/Hypothesis:
Several recent studies have observed a high incidence of duodenoscope microbial contamination and an association of contamination with healthcare-acquired infections. This study sought to quantify nasopharyngoscope microbial contamination relative to that of other endoscope categories and characterize the manufacturers, outcomes, and microbial profiles associated with these cases.
Study Design:
Retrospective, cross-sectional study.
Methods:
A total of 3,865 adverse events were collected from 2013 to 2019 using the US Food and Drug Administration Manufacturer and User Facility Device Experience database. The fraction of total device failures associated with contamination was quantified for nasopharyngoscopes, bronchoscopes, duodenoscopes, and gastroscopes. Odds ratios of nasopharyngoscope contamination compared to that of bronchoscopes, duodenoscopes, or gastroscopes were calculated, and significance was assessed by χ2 analysis. The Kruskal-Wallis test was used for nonparametric testing of significance.
Results:
Nasopharyngoscope device failures were reported at an incidence of 0.646 per month; 34.1% involved contamination, comparable to the frequency observed for bronchoscopes (23.4%, P = .118), duodenoscopes (29.2%, P = .493), and gastroscopes (45.3%, P = .178). The frequency of device contamination was observed to be significantly higher for a particular endoscope manufacturer regardless of endoscope category (Kruskal-Wallis P = .021). In instances of contamination, nasopharyngoscopes were significantly less associated with patient harm or death than bronchoscope (odds ratio [OR] = 10.2) and duodenoscope (OR = 4.81) cases.
Conclusions:
Although the rates of contamination were comparable across all endoscope categories, nasopharyngoscope contamination was less commonly associated with patient harm or death. In an era of rising healthcare costs, determining adequate disinfection standards for nasopharyngoscopes and their impact on patient safety is crucial.
Level Of Evidence:
NA Laryngoscope, 131:598-605, 2021.
Insights
Nasopharyngoscope contamination rates are similar to other endoscopes, but less likely to cause patient harm. Ensuring proper disinfection is vital for patient safety and managing healthcare costs.
Area of Science:
- Medical device safety
- Infection control
- Gastroenterology and Pulmonology
Background:
- Recent studies highlight high duodenoscope microbial contamination and links to healthcare-acquired infections.
- Concerns exist regarding microbial contamination across various endoscope types.
Purpose of the Study:
- To quantify nasopharyngoscope microbial contamination compared to other endoscopes.
- To identify manufacturers, outcomes, and microbial profiles associated with endoscope contamination.
Main Methods:
- Retrospective analysis of 3,865 adverse events from the FDA MAUDE database (2013-2019).
- Quantified contamination rates for nasopharyngoscopes, bronchoscopes, duodenoscopes, and gastroscopes.
- Calculated odds ratios for contamination and patient harm, using statistical analysis (χ², Kruskal-Wallis).
Main Results:
- Nasopharyngoscope contamination incidence was 0.646/month, with 34.1% involving contamination.
- Contamination frequencies were comparable across nasopharyngoscopes (34.1%), bronchoscopes (23.4%), duodenoscopes (29.2%), and gastroscopes (45.3%).
- One manufacturer showed significantly higher contamination rates across endoscope types (P=.021).
- Nasopharyngoscope contamination was less associated with patient harm (OR=4.81-10.2) than other endoscopes.
Conclusions:
- Despite comparable contamination rates, nasopharyngoscope contamination poses a lower risk of patient harm.
- Adequate disinfection standards for nasopharyngoscopes are crucial for patient safety and cost-effective healthcare.
Related Concept Videos
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Studies I: Bronchoscopy and Thoracoscopy
Bronchoscopy
Description
Bronchoscopy is a procedure that involves direct visualization of the larynx, trachea, and bronchi for diagnostic and therapeutic purposes. A flexible fiber optic or rigid bronchoscope is used to carry out the procedure. The fiber-optic bronchoscope is more frequently used due...
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures II: Colonoscopy