Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Is the strain pattern of conventional stems negatively affected by a previously short stem THA? An experimental study
Thilo Floerkemeier1,2, Romy Arokiasamy1, Stefan Budde1
1Department of Orthopaedic Surgery, Hannover Medical School, Hannover, Germany.
Background:
A short stem hip arthroplasty can be revised in many cases using a conventional stem. Furthermore, in some cases the implantation of a short stem is intended, but intraoperatively reasons may lead to the decision to implant a conventional stem after previous preparation of a short stem.
Objective:
In both cases it is questionable if the anchorage of a conventional stem is negatively affected by the previous preparation of a short stem. In clinical practice mid- or long-term follow up for these special cases hardly exist.
Methods:
The strain patterns for the conventional Bicontact stem in primary implantation and after preparation of the proximal femur for a METHA short stem were tested biomechanically in three pairs of cadaveric femora.
Results:
The strain patterns for the conventional Bicontact after preparation of the METHA short stem were similar to conditions after testing the conventional stem in primary conditions.
Conclusions:
These data lead to the consequence that in clinical practise the implantation of a conventional stem after preparation of a short stem and even after revision of a short stem is possible without increased risk of loosening or long-term stress-shielding.