Clothing Effects on Limb Tourniquet Application
Related Experiment Video
Updated: Dec 17, 2025

Remote Limb Ischemic Preconditioning: A Neuroprotective Technique in Rodents
Published on: June 2, 2015
Background:
Sometimes tourniquets are applied over clothing. This study explored clothing effects on pressures and application process.
Methods:
Generation 7 Combat Application Tourniquets (C-A-T7), Generation 3 SOF® Tactical Tourniquets-Wide (SOFTTW), Tactical Ratcheting Medical Tourniquets (Tac RMT), and Stretch Wrap And Tuck Tourniquets (SWATT) were used with different clothing conditions (Bare, Scrubs, Uniform, Tights) mid-thigh and on models (ballistic gel and yoga mats).
Results:
Clothing affected pressure responses to controlled force applications (weight hangs, n=5 thighs and models, nonlinear curve fitting, p < .05). On models, clothing affected secured pressures by altering surface interactions (medians: Gel Bare C-A-T7 247mmHg, SOFTTW 99mmHg, Tac RMT 101mmHg versus Gel Clothing C-A-T7 331mmHg, SOFTTW 170mmHg, Tac RMT 148mmHg; Mats Bare C-A-T7 246mmHg, SOFTTW 121mmHg, Tac RMT 99mmHg versus Mats Clothing C-A-T7 278mmHg, SOFTTW 145mmHg, Tac RMT 138mmHg). On thighs, clothing did not significantly influence secured pressures (n=15 kneeling appliers, n=15 standing appliers) or occlusion and completion pressures (n=15). Eleven of 15 appliers reported securing on clothing as most difficult. Fourteen of 15 reported complete applications on clothing as most difficult.
Conclusions:
Clothing will not necessarily affect tourniquet pressures. Surface to tourniquet interactions affect the ease of strap sliding, so concern should still exist as to whether applications over clothing are dislodged in a distal direction more easily than applications on skin.
Related Concept Videos
Peripheral Artery Disease V: Postoperative Nursing Management
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Pre-Procedural Guidelines for Assessing Blood Pressure
Venous Thrombosis IV: Nursing Management
Arteries of Lower Limbs