Comparison of quantitative image quality of cardiac computed tomography between raw-data-based and model-based
1Asan Medical Center, Department of Radiology and Research Institute of Radiology, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul, 05505, South Korea.
Background:
Image sharpness is commonly degraded on cardiac CT images reconstructed using iterative reconstruction algorithms.
Objective:
To compare the image quality of cardiac CT between raw-data-based and model-based iterative reconstruction algorithms developed by the same CT vendor in children and young adults with congenital heart disease.
Materials And Methods:
In 29 patients with congenital heart disease, we reconstructed 39 cardiac CT datasets using raw-data-based and model-based iterative reconstruction algorithms. We performed quantitative analysis of image sharpness using distance25-75% and angle25-75% on a line density profile across an edge of the descending thoracic aorta in addition to CT attenuation, image noise, signal-to-noise ratio and contrast-to-noise ratio. We compared these quantitative image-quality measures between the two algorithms.
Results:
CT attenuation did not show significant differences between the algorithms (P>0.05) except in the aorta. Image noise was small but significantly higher in the model-based algorithm than in the raw-data-based algorithm (4.8±2.3 Hounsfield units [HU] vs. 4.7±2.1 HU, P<0.014). Signal-to-noise ratio (110.2±50.9 vs. 108.4±50.4, P=0.050) and contrast-to-noise ratio (91.0±45.7 vs. 89.6±45.1, P=0.063) showed marginal significance between the two algorithms. The model-based algorithm showed a significantly smaller distance25-75% (1.4±0.4 mm vs. 1.6±0.3 mm, P<0.001) and a significantly higher angle25-75% (77.0±4.3° vs. 74.1±5.7°, P<0.001) than the raw-data-based algorithm.
Conclusion:
Compared with the raw-data-based algorithm, the model-based iterative reconstruction algorithm demonstrated better image sharpness and higher image noise on cardiac CT in patients with congenital heart disease.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...