Determining validity of the PALite and ODFS PACE activity logger for measuring step count in healthy adults
L Miller Renfrew1, D Rafferty2, A Lord1
1MS Service, NHS Ayrshire and Arran, Scotland, UK.
Background:
Determining adherence with orthoses is important for clinicians prescribing devices. Measuring orthotic use often relies on patient recall which has poor agreement with objective measures. Measuring step count whilst wearing an orthosis could help objectively quantifying adherence. The Odstock Drop Foot Stimulator (ODFS) Pace, used in foot drop, has an integral activity logger which provides data on step count. The PALite, an accelerometer, measures step count and can be fixed to an ankle foot orthoses (AFO). Both have the potential to provide objective measures of adherence; however, their validity for this purpose has not been determined.
Research Question:
To determine the validity of the PALite and ODFS Pace activity logger in measuring total step count, by exploring their level of agreement.
Methods:
A convenience sample of sixteen healthy volunteers, aged 18-65, were recruited from Glasgow Caledonian University (GCU). Participants walked continuously for 5 min on a treadmill at three walking speeds; normal (1.3ms-1), slow (0.4 ms-1) and fast (1.7-2.0 ms-1), wearing both the PALite and ODFS Pace. All walks were video recorded, viewed by 2 raters, and observed step count was determined by a click counter. Step count from both devices was validated against observed step count using video recording. The level of agreement between the three methods was established.
Results:
There was no significant difference between the 3 methods of measuring step count at any walking speed (normal, p = 0.913; slow, p = 0.938; fast, p = 0.566). Good levels of agreement for both devices with observed step count at all 3 walking speeds, with mean percentage differences being between -1.2 and 2.1 % (maximum upper and lower levels of agreement = 19.5 and -18.8 %) was detected.
Significance:
Clinicians could consider both devices to objectively measure step count with people who are prescribed foot drop orthoses, thus quantifying orthotic use.
More Related Videos
Related Concept Videos
Special considerations while measuring pulse
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of blood pressure in brachial artery(two-step method)
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure in the Leg
Preparation: