Related Experiment Videos
Long-term patency of polytetrafluoroethylene graft fistulas
E S Schuman1, G F Gross, J F Hayes
1Department of Surgery, Good Samaritan Hospital and Medical Center, Portland, Oregon.
Abstract:
Polytetrafluoroethylene grafts have been the mainstay for hemoaccess at our dialysis unit since 1976. Of 528 grafts placed in the forearm, 81 were evaluated for long-term patency and complication rates. When appropriate diagnostic studies and clinical signs delineate the cause of graft problems, a variety of surgical techniques can be used to maintain graft patency. The overall patency rate was 81 percent at 4.47 years. There were 3.5 complications per graft and 1 procedure per 1.25 years of follow-up. Over two thirds of the grafts had minimal complications and showed a 95 percent patency rate at 4.76 years. Impra grafts seemed to fare better than Gore-tex grafts, but their patency and duration were similar. Polytetrafluoroethylene grafts are easy to use, easy to revise and repair, have a long shelf life, and can withstand repeated dialysis access for many years.