High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
First-Line Toxicological Screening with Fully Automated Extraction
Tiphaine Robin1, Souleiman El Balkhi1, Sylvain Dulaurent1
1Department of Pharmacology, Toxicology and Pharmacovigilance, Limoges University Hospital, Rue du Pr Descottes, 87270 Limoges, France.
A new automated screening method identifies and quantifies 245 compounds in clinical toxicology. This rapid LC-MS-MS procedure offers high confidence for drug overdose and poisoning detection in patient samples.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
- Forensic Toxicology
Background:
- Clinical toxicology laboratories require rapid, unambiguous methods for identifying compounds in overdose or poisoning cases.
- Existing methods may lack the speed and comprehensive identification needed for timely clinical decisions.
- The demand for high-throughput screening in toxicology necessitates advanced analytical techniques.
Purpose of the Study:
- To develop and validate a fully automated screening procedure for identifying and quantifying 245 clinically relevant compounds.
- To achieve unambiguous compound identification and reliable quantification in a single analytical run.
- To provide a rapid and high-confidence screening method for clinical toxicology laboratories.
Main Methods:
- Development of a fully automated sample preparation using a programmable liquid handler.
- Coupling of automated extraction with liquid chromatography-tandem mass spectrometry (LC-MS-MS).
- Utilized positive/negative ionization modes with up to 15 multiple reaction monitoring (MRM) transitions per compound for identification and quantification.
Main Results:
- The method successfully identified and quantified 245 compounds with high confidence using 15 MRM transitions per compound.
- Validated quantitative performance showed satisfactory intra- and inter-assay imprecision and inaccuracy (CV% < 20%).
- Excellent correlation (R² = 0.96) was observed when compared to dedicated quantitative methods using 175 sample pairs.
Conclusions:
- The developed automated LC-MS-MS screening procedure provides high confidence in compound detection for clinical toxicology.
- The method's speed (8 min sample prep + 10 min chromatography) and reliability make it suitable for core laboratories.
- This first-line screening approach aids in rapid and accurate diagnosis of drug overdose and poisoning.