Louis-Marie Galerneau1,2, Nicolas Terzi1,2, Emanuele Turbil3
1Médecine Intensive Réanimation, Université de Grenoble-Alpes, Grenoble, France.
Background:
Automatic tube compensation (ATC) unloads endotracheal tube (ETT) resistance. We conducted a bench assessment of ATC functionality in ICU ventilators to improve clinical management.
Methods:
This study had 2 phases. First, we performed an international survey on the use of ATC in clinical practice, hypothesizing a rate of ATC use of 25%. Second, we tested 7 modern ICU ventilators in a lung model mimicking a normal subject (Normal), a subject with ARDS, and a subject with COPD. Inspiratory effort consisted of esophageal pressure over 30 consecutive breaths obtained in a real patient under weaning. A brand new 8-mm inner diameter ETT was attached to the lung model, and ATC was set at 100% compensation for the ETT. The 30 breaths were first run with ATC off and no ETT (ie, reference period), and then with ATC on and ETT (ie, active period). The primary end point was the difference in tidal volume (VT) between reference and active periods. We hypothesized that the VT difference should be equal to 0 in an ideally functioning ATC. VT difference was compared across ventilators and respiratory mechanics conditions using a linear mixed-effects model.
Results:
The clinical use of ATC was 64% according to 644 individuals who responded to the international survey. The VT difference varied significantly across ventilators in all respiratory mechanics configurations. The divergence between VT difference and 0 was small but significant: the extreme median (interquartile range) values were -0.013 L (-0.019 to -0.002) in the COPD model and 0.056 L (0.051-0.06) in the Normal model. VT difference for all ventilators was 0.015 L (95% CI 0.013-0.018) in the ARDS model, which was significantly different from 0.021 L (95% CI 0.018-0.024) in the Normal model (P < .001) and 0.010 L (0.007-0.012) in the COPD model (P = .003).
Conclusions:
ATC is used more frequently in clinical practice than expected. In addition, VT delivery by ATC differed slightly though significantly between ventilators.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Cardiopulmonary Resuscitation II: ACLS Airway Management