Comparison of two digital intraoral radiography imaging systems as a function of contrast resolution and exposure
Kader C Aydin1, Oğuzhan Demirel2, Mutlu Özcan3
1Department of Dentomaxillofacial Radiology, School of Dentistry, Istanbul Medipol University, Istanbul, Turkey - kadercesur@yahoo.com.
Background:
To compare the image quality of two different digital imaging systems; one photostimulable phosphor plate system (PSP) and a direct digital radiography system with CMOS imaging sensor; via evaluating contrast resolution among four different exposure times.
Methods:
Endodontically treated incisor teeth embedded in paraffin blocks are aligned next to a 99.5% Al wedge and exposed for 0.8, 0.1,0.125 and 0.16 seconds using both the CMOS and PSP systems. Using ImageJ software, 5 isometric and isogridded ROI from each root filling area and isometric ROI from the Al stepwedge were calculated.
Results:
Evaluation of the total of 120 images displayed that PSP system produced significantly higher contrast resolution (P<0.05) in regard to pixel values than the CMOS. The CMOS system was non- responsive to increasing dose (P=0.000). Regarding the EqAl values, no significant difference was determined between groups (P>0.05).
Conclusions:
The contrast resolution was higher using the PSP system. It can be estimated that, filling material will be more obvious under lower doses using PSP.
Related Concept Videos
Imaging Studies III: Computed Tomography
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...