What is the mechanism of irregular pacing spikes? Pacing failure, sensing failure, or something else? A case report
Gaku Oguri1,2, Toshiya Kojima1, Katsuhito Fujiu1,3
1Department of Cardiovascular Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo, Tokyo 113-8655, Japan.
Background:
In cardiac implantable electronic device management, confused electrocardiograms are sometimes encountered.
Case Summary:
We experienced a case of dizziness and presyncope, which was accompanied by irregular pacing spikes following QRS complexes, pacing failure, sensing failure, or something else. The patient's second active pacemaker in her left chest was functioning normally. However, her old abdominal pacemaker had automatically changed from the OVO to VVI mode because of an elective replacement indicator. As a result, her old and second pacemakers had an interaction. Old invalid pacing spikes induced oversensing of the current pacemaker, which interfered with valid pacing.
Discussion:
We experienced a rare case of dizziness and presyncope with confused pacing spikes. Physicians should be aware of the unexpected interactions that are possible with devices with automatic mode changes.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias I: Introduction
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias IV: Characteristics of Bradyarrhythmias