Relation between age and CT radiation doses: Dose trends in 705 pediatric head CT
Related Experiment Video
Updated: Dec 16, 2025

Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Mohammad Hassan Kharita1, Huda Al-Naemi1, Chiara Arru2
1Hamad Medical Corporation, Doha, Qatar.
Insights
Pediatric head CT radiation doses increase linearly with patient age. While most CT dose index volume (CTDIvol) levels were below European reference levels, some dose length products (DLP) exceeded them, highlighting the need for updated technologist guidelines.
Area of Science:
- Radiology
- Medical Physics
- Pediatric Imaging
Background:
- Pediatric head CT scans are common diagnostic procedures.
- Automatic tube potential selection and tube current modulation are used to optimize radiation dose.
- Understanding the relationship between patient age and radiation dose is crucial for pediatric radiation safety.
Purpose of the Study:
- To investigate the correlation between patient age and radiation doses (CTDIvol and DLP) in pediatric head CT.
- To evaluate the effectiveness of automatic exposure control techniques in managing pediatric radiation doses.
Main Methods:
- Retrospective analysis of 705 pediatric head CT scans.
- Data collected included patient demographics, scan parameters, CTDIvol, and DLP.
- Scans were performed using multidetector-row CT with automatic tube potential selection (Care kV) and tube current modulation (Care Dose 4D).
- Statistical analysis included linear correlation and ANOVA.
Main Results:
- A significant linear correlation was found between radiation doses and patient age (r² = 0.66, p < 0.001).
- Median CTDIvol was below European Diagnostic Reference Levels (EDRL) for most patients.
- Median DLP exceeded EDRL for 28% of patients, primarily due to longer scan lengths.
Conclusions:
- Radiation dose in pediatric head CT increases linearly with patient age.
- While CTDIvol is generally well-controlled, DLP can exceed reference levels, necessitating careful protocol management.
- Clearer guidelines are needed for technologists to optimize radiation dose in pediatric head CT examinations.
Purpose:
To evaluate the relationship between patient age and radiation doses associated with routine pediatric head CT performed with automatic tube potential selection and tube current modulation techniques.
Methods:
We obtained patient demographics, scan parameters, and radiation dose descriptors (CT dose index volume -CTDIvol and dose length product -DLP) associated with consecutive routine head CT in 705 children (mean age 6.9 ± 5 years). Children were scanned on one of the three multidetector-row CTs (64-128 slices, Siemens) over 6 months period in a tertiary hospital. All head CT exams were performed in helical scan mode using automatic tube potential selection (Care kV) and automatic tube current modulation (Care Dose 4D) techniques. The information was obtained from a radiation dose monitoring software. Data were analyzed using linear correlation and analysis of variance.
Results:
Most age-wise median CTDIvol (9-27 mGy; 703/705 pediatric head CT, >99 %) from our institution were lower than the European Diagnostic Reference Levels (EDRL, CTDIvol 24-50 mGy) but median DLP (151-586 mGy cm) from 201/705 children (28 %) was higher than the EDRL (DLP 300-650 mGy cm). Unlike the age-stratified EDRL, a combination of automatic tube potential selection and tube current modulation for pediatric head results in a significant linear correlation between radiation doses and patient age (r2 = 0.66, p < 0.001).
Conclusions:
Radiation doses for head CT change linearly with children's age. Despite lower CTDIvol and DLP for most children, longer scan length resulted in higher DLP for some pediatric head CT compared to the corresponding EDRL; this result underscores the need to promote clear guidelines for technologists operating CT.
Related Concept Videos
Drug Dosing: Infants and Children
Pharmacokinetics in Pediatric Patients: Drug Distribution
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Pharmacokinetics in Pediatric Patients: Drug Metabolism
Pharmacokinetics in Pediatric Patients: Drug Excretion