Movement during focused ultrasound therapy caused by an unstable magnetic resonance table: case report
Kazuaki Yamamoto1,2, Hisashi Ito3, Shigeru Fukutake3
11Department of Neurosurgery, Tokyo Women's Medical University, Tokyo.
Abstract:
Transcranial MR-guided focused ultrasound (MRgFUS) therapy is a less invasive form of stereotactic treatment for tremors and other movement disorders. Its stereotactic accuracy is ensured by stability of the stereotactic frame and MR table. The authors report a case wherein the patient's movement was detected, and the MR images were repeated to continue the treatment. A 72-year-old man with essential tremor underwent unilateral ventralis intermedius thalamotomy using MRgFUS. The stereotactic frame was correctly fixed to the patient's skull and the table. During the seventh sonication, the patient pressed the emergency button and vomited several times. Before the eighth sonication, the patient's movement was detected and was verified on coronal images. The MR images were repeated, and the treatment was successfully completed with significant improvement in the tremors. After treatment, it was discovered that the MR table was laterally unstable due to the absence of ball bearings, which should be present on both sides of the guide rail of the MR table. The ball bearings were attached to the reverse side of the table, and the table was stabilized. Stereotactic accuracy of MRgFUS is not only ensured by rigid fixation of the stereotactic frame, but also by stability of the MR table.
More Related Videos
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
09:07Sonodynamic Therapy for the Treatment of Glioblastoma Multiforme in a Mouse Model Using a Portable Benchtop Focused Ultrasound System
Published on: February 10, 2023