Initial multicenter experience with a new high-density coloring module: impact for complex atrial arrhythmias
Grégoire Massoullié1, Ghassan Moubarak2, Olivier Thomas2
1Cardiology Department, CHU Clermont-Ferrand, Clermont University, ISIT-CaVITI, 58 rue Montalembert, 63000 CHU, Clermont-Ferrand, France. gmassoullie@chu-clermontferrand.fr.
Background:
High-density automated mapping of complex atrial tachycardias (ATs) requires accurate assessment of activation maps. A new local activation display module (HD coloring, Biosense Webster®) provides higher map resolution, a better delineation of potential block reducing color interpolation, and a new propagation display. We evaluated the accuracy of a dedicated local activation display compared with standard algorithm.
Methods:
High-density maps from 10 AT were collected with a multipolar catheter and were displayed with standard activation or HD coloring. Six expert operators retrospectively analyzed activation maps and were asked to define (1) the tachycardia mechanism, (2) ablation target, and (3) level of difficulty to interpret those maps.
Results:
Using HD coloring, operators were able to reach a correct diagnosis in 93% vs. 63%, p < 0.05 compared to standard activation maps. Time to diagnosis was shorter 1.9 ± 1.0 min vs. 3.9 ± 2.1 min, p < 0.05. Confidence level would have allowed ablation without necessity for entrainment maneuvers in 87% vs. 53%, p < 0.05. Operators would have needed to remap or proceed with multiple entrainments in 3% vs. 13% of cases, p < 0.05. Finally, ablation strategy was more accurately identified in 97% vs. 67%, p < 0.05.
Conclusion:
Activation mapping with the new HD coloring module allowed a more accurate, reliable, and faster interpretation of complex ATs mechanisms compared to standard activation maps.
More Related Videos
09:17High-Resolution Endocardial and Epicardial Optical Mapping in a Sheep Model of Stretch-Induced Atrial Fibrillation
Published on: July 29, 2011
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization