X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Evaluation of a low-dose neonatal chest radiographic system
E M Burton1, D R Kirks, J L Strife
1Department of Radiology, Children's Hospital Medical Center, Cincinnati, OH 45229.
Abstract:
A new low-dose chest radiographic system for use in the neonatal nursery was evaluated. This test system, composed of a Du Pont Kevlar fiber-front cassette, Quanta fast-detail screen, Cronex 4L film (wide latitude), and additional yttrium filtration (0.1 mm), reduced the radiation dose in neonatal chest radiography by 69% (0.9 vs 2.9 mrad [0.009 vs 0.029 mGy]) as compared with a conventional system without added yttrium filtration; the thyroid dose was reduced by 76% (0.9 vs 3.7 mrad [0.009 vs 0.037 mGy]). The cumulative dose reduction was achieved through a combination of factors, including (1) beam hardening by the added yttrium filter, (2) increased X-ray transmission through the Kevlar cassette, and (3) a fast film-screen combination. Scatter radiation at distances of 1 and 6 ft. (0.3 and 1.8 m) was negligible for both systems. Image sharpness was compared for the conventional system with and without added yttrium filtration and for the Kevlar system with yttrium. Although sharpness of bony detail was unchanged by adding yttrium filtration to the conventional system, a decrease in sharpness was noted with the Kevlar system. Because image sharpness was affected in the test system, we are not using the Kevlar-Cronex 4L system for mobile chest radiography in the neonatal intensive care unit, despite dose reductions. However, further study is recommended to determine if there is a slower film-screen combination with yttrium filtration that will not degrade image sharpness.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...