Diagnostic reference levels for fluoroscopically guided procedures in a South African tertiary hospital
Leon Malan1, Richard D Pitcher1, Michelle da Silva1
1Division of Radiodiagnosis, 71942Tygerberg Hospital, University of Stellenbosch, Cape Town, Western Cape, South Africa.
Background:
The burgeoning usage and complexity of fluoroscopically guided procedures (FGPs) contribute to extended examination times and increased risk of adverse radiation effects. Diagnostic reference levels (DRLs) play a pivotal role in dose optimization. There are limited DRL data for FGPs in low- and middle-income countries (LMICs).
Purpose:
To determine local DRLs (LDRLs) for common FGPs in the South African (SA) context and compare these with published international data.
Material And Methods:
A three-year, retrospective study of the 15 most frequently performed FGPs at a SA institution. For each procedure, the 50th and 75th percentiles of kerma area product (KAP), reference point air kerma (Ka,r), and fluoroscopy time data were derived. Published international FGP DRL data were collated and compared with the 75th percentiles of local institutional dosage parameters.
Results:
The commonest FGPs were aorto-bifemoral diagnostic angiography (n = 590), aorto-bifemoral interventional angiography (n = 287), nephrostomy (n = 265), and bronchial arterial embolization (BAE) (n = 208). Selective abdominal vessel interventional angiography (KAP = 170 Gy . cm2; Ka,r = 877 mGy) recorded the highest LDRL dosages; BAE was the longest procedure (LDRL = 38 min). Nephrostomies achieved the lowest LDRLs across all parameters (KAP = 10 Gy . cm2; Ka,r = 63 mGy, fluoroscopy time = 4.3 min). All Tygerberg Hospital LDRLs with comprehensive comparable data were within or below published ranges.
Conclusion:
This study advances international radiation protection initiatives, addresses the paucity of LMIC DRL data, demonstrates broad alignment of Tygerberg Hospital FGP practice with international norms and highlights areas for optimization of institutional practice.
More Related Videos
09:03Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...