Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Characterisation of a hand-held CZT-based gamma camera for 177Lu imaging
Daniel Roth1, Erik Larsson2, Anna Sundlöv3,4,5
1Department of Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden. daniel.roth@med.lu.se.
Background:
Currently, hand-held gamma cameras are being developed for 99mTc imaging, mainly for sentinel lymph node detection. These cameras offer advantages, such as mobility and ease of access, and may be useful also for other applications such as biokinetic studies in animals or for imaging of small, superficial structures in patients. In this work, the suitability of a CZT-based hand-held camera for 177Lu imaging is investigated. The energy response of CZT-based detectors combined with the multiple photon emissions of 177Lu poses new challenges compared to 99mTc imaging, and a thorough camera characterisation is thus warranted.
Methods:
Three collimators (LEHR, LEHS, and MEGP) and three energy windows (55 keV, 113 keV, and 208 keV) are investigated. Characterised camera properties include the system spatial resolution, energy resolution, sensitivity, image uniformity, septal penetration, and temperature dependence. Characterisations are made starting from NEMA guidelines when applicable, with adjustments made when required. The applicability of the camera is demonstrated by imaging of a superficially located tumour in a patient undergoing [177 Lu]Lu-DOTA-TATE therapy.
Results:
Overall, the results are encouraging. Compared to a conventional gamma camera, the hand-held camera generally has a higher sensitivity for a given collimator. For source-collimator distances below 3 cm, the spatial resolution FWHM is within 6 mm for the LEHR and MEGP collimators. Before uniformity correction, the central field-of-view integral uniformity shows best results for the 113-keV window, with values obtained between 11 and 14%. The corresponding values after uniformity correction are within 3%. Effects of septal penetration are observed but are manageable with a proper combination of collimator and energy window setting. Septal penetration and collimator scatter not only affect the 208-keV window but also contribute with counts in lower windows due to energy-tailing effects. The patient study revealed non-uniform uptake patterns in a region that appeared uniform in a conventional gamma camera image.
Conclusions:
The results show that the hand-held camera can be used for 177Lu imaging. A 113-keV energy window combined with LEHR or MEGP collimators provides the best image system characteristics.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...